Strona 1 - Owner’s Booklet

Owner’s BookletComplete Diabetes Monitoring SystemFor diabetes and life.

Strona 2 - PLUS OWNER:

3GETTING STARTEDThe ONE TOUCH™BASIC™PlusDiabetes Monitoring Systemconsists of three main products: theONE TOUCH™BASIC™Plus BloodGlucose Meter, ONE TOU

Strona 3 - Plus Diabetes Monitoring

4ONE TOUCH™BASIC™Plus Meterhas a number of settings you canchange, including the language, timeand date, display formats, and theunit of measure. For

Strona 4

ONE TOUCH™BASIC™PlusBlood Glucose MeterStudy this diagram and becomefamiliar with all the parts of yourONE TOUCH™BASIC™Plus Meter.5CMCMON/OFF BUTTONTh

Strona 6

7TESTSPOTOR THISNOTE: Your test strip may look like THISONE TOUCH™Test StripsThe ONE TOUCH™BASIC™PlusDiabetes Monitoring Systemmeasures the amount of

Strona 7

8IMPORTANT TEST STRIP INFORMATION ONE TOUCH™Test Strips can be damaged by heat and light. Keep themsealed in the original vial. Store the vial in a co

Strona 8 - LIMITATIONS

9CAUTION: If the code number onthe meter display does not matchthe code number on the vial ofONE TOUCH Test Strips, testresults may be false.BEFORE TE

Strona 9 - Infection Control

10Step 1: Press the On/Off Button.When turning meter on, all displaysymbols will appear. The last testresult will then appear on the display for a mom

Strona 10 - GETTING STARTED

11Step 3: Code the Meter.While the meter displays code--, or after the word insert is displayed,press and release the C button; thecode number will ap

Strona 11

12(Example)Checking with the Check StripA purple and white check strip isincluded in the carrying case ofyour ONE TOUCH™BASIC™PlusSystem. Before doing

Strona 12 - Blood Glucose Meter

DEAR ONE TOUCHTMBASICTMPLUS OWNER:Congratulations! You’ve chosen a very easy-to-use blood glucose (sugar) monitoringsystem for home testing. When used

Strona 13

How to do a check strip test:Step 1: Press the On/Off Button.Insert Side 1 of the Check Strip.When insert code 10 (Example)strip appears, slidethe not

Strona 14 - Test Strips

14If these messages appear, repeat thecheck strip test. If these messagesappear again, clean your meter andrepeat the test. If you do not obtainan acc

Strona 15 - Test Strip insert

15 Do not scratch the check strip. Do not leave the check strip insunlight. If you lose or damage the checkstrip, call your authorized LifeScanreprese

Strona 16 - BEFORE TESTING

16Do a control solution test: For practice, before testing yourblood with the ONE TOUCH™BASIC™PlusMeter for the firsttime When you begin using a new v

Strona 17 - (Example)

17*For detailed information on ONE TOUCH™Control Solution, refer to the ONE TOUCH™Control Solution insert.IMPORTANT CONTROL SOLUTION TEST INFORMATION

Strona 18 - Checking the System

18How to do a control solution test:Step 1: Press On/Off Button. InsertTest Strip.The code number thatappears on the meterdisplay should matchthe code

Strona 19

Step 2: Apply Control Solution.Shake the controlsolution vial wellbefore using. If thereare any bubbles onthe top of the vial,wipe them off with aclea

Strona 20

20Control solution test results areautomatically stored with the wordcontrl. (If you use a logbook,indicate control solution tests by writing “C” afte

Strona 21

CM21Detailed Test ProcedureChoose a clean, dry work surface. Make sure you have all the items needed to test: ONE TOUCH™BASIC™PlusMeter ONE TOUCH™Test

Strona 22

22Getting a Drop of BloodCAUTION: To reduce the chance of infection: Never share a lancet with another person. Always use a new, sterileLifeScan Lance

Strona 23

The ONE TOUCH™BASIC™Plus Diabetes MonitoringSystem is intended for in vitro diagnostic use forquantitative determination of glucose in whole bloodas a

Strona 24 - Control Solution insert

While the lancet is in the Penlet™Plus Sampler, twist off theprotective disk. Make two full turnsto ensure that the lancet separatesfrom the protectiv

Strona 25

24Step 2: Cock the Penlet™PlusSampler.Slide the ejection/cocking controlbackward until it clicks. If it does notclick, the Penlet Plus Sampler mayhave

Strona 26

25Step 4: Lance Your Finger.Choose a spot on the side of adifferent finger each time you test.Repeated punctures in the same spot may cause soreness a

Strona 27

26Optional: To replace the protectivedisk on the lancet, place the disk ona hard surface and push the exposedtip into the protective disk.Always use c

Strona 28 - Detailed Test Procedure

How to Test Your BloodStep 1: Press On/Off Button. MakeSure Code Number Displayed onMeter Matches Code Number onTest Strip Vial. Insert Test Strip.Mak

Strona 29

28Remove a test stripfrom the vial. Replacethe vial cap immedi-ately. Do not touch thetest spot on the teststrip. The test spotshould be white orivory

Strona 30

29Step 2: Apply Blood Sample.With the test strip in the meter andon the display, apply a drop of bloodto the test spot.Be sure that you: Touch only th

Strona 31

30Step 3: Accurate Results in 45 Seconds.The meter then countsdown from 45 to 0seconds, beeps (if theBeep option is set to ), and displays yourresult.

Strona 32

31Special MessagesThe meter displays results from 0 to600 mg/dL (33.3 mmol/L). If yourblood glucose test result is 60 mg/dL(3.3 mmol/L) or below, your

Strona 33

32Checking the Amount ofBlood on the Test StripAlthough the ONE TOUCH™BASIC™Plus System requires only a small drop of blood, it is veryimportant that

Strona 34

CAUTION: Before using any product to test yourblood glucose, read all instructions and practicethe test. Do all quality control checks as directedand

Strona 35

33If the following messageappears on the display, your drop ofblood was too small, or smeared, orthe test strip was not pushed all the way into the te

Strona 36

Look at the back of the test strip.You should see a full, dark circle.If there are any white patches orstreaks, you may have smearedthe blood or the

Strona 37

35Entering the Memory ModeTurn the meter on by pressing theon/off button. You can access themeter’s memory and recall testresults at any time before o

Strona 38 - Special Messages

36Recalling Test ResultsPressing the M button allows you tosee each blood glucose, check strip,and control solution test stored inmemory. The records

Strona 39 - Blood on the Test Strip

37SETTING METEROPTIONSYou do not have to enter the set-upmode to test. However, you cancustomize the various option settingsof your ONE TOUCH™BASIC™Pl

Strona 40 - (Correct)

38 Use the M button to scroll throughthe options or move within anoption. If you want to change an optionsetting, use the C button.* Time and Date opt

Strona 41 - USING THE METER

39Clock (Time and Date)The ONE TOUCH™BASIC™PlusMeter has an internal clock that willrecord the time and date of eachtest you perform. Your meter hasbe

Strona 42 - Entering the Memory Mode

40Use the C button to change the min-utes. Press the M button to see thenew time setting (the colon willflash). Press the M button again to move tothe

Strona 43 - Recalling Test Results

41Press the M button and the day willflash. Press the C button to change theday. Pressing the M button again willdisplay the new date setting.You may

Strona 44 - SETTING METER

42LanguageThe ONE TOUCH™BASIC™PlusMeter can display messages in 17 languages: English [engl.], Spanish [espan.],French [franc.], Italian [italia],Hung

Strona 45 - Clock option is on

TABLE OF CONTENTSPRECAUTIONS AND LIMITATIONS 1Infection Control 2 GETTING STARTED 3The Complete ONE TOUCH™BASIC™PlusDiabetes Monitoring System 3ONE TO

Strona 46 - Clock (Time and Date)

43Time FormatThe meter can be set to display thetime in either a 12-hour (AM/PM)format or a 24-hour format.To change the time format, enter the set-up

Strona 47

44Unit of MeasureThe meter is set to display testresults in milligrams per deciliter(mg/dL). You have the option tochange to millimoles per liter(mmol

Strona 48 - Beep Prompts

45To change the decimal separator,enter the set-up mode (see page 37)and select mmol/L as the unit ofmeasure. Use the M button to choosethe Decimal Se

Strona 49 - Language

46SOLVING PROBLEMSThe messages that appear on your ONE TOUCH™BASIC™Plus Meterdisplay guide you through the testing procedure, alert you to problems as

Strona 50 - Date Format

47The meter is turned on. This is a reminder thatprior to testing you should verify that the codedisplayed on the meter matches the code on thetest st

Strona 51 - Decimal Separator

48A blood test has been completed. This indicates ablood glucose result when the meter is set for mg/dL.A blood test has been completed. This indicate

Strona 52 - Exiting the Set-Up Mode

The check strip test result falls into the acceptable range. The check strip range is printed on the back ofyour meter.Your blood glucose test result

Strona 53 - SOLVING PROBLEMS

50MESSAGE PROBLEM WHAT TO DOSome parts of the dis-play are not working.The messages will beincomplete.Call your authorizedLifeScan representative.Erro

Strona 54 - MESSAGE APPEARS WHEN

51MESSAGE PROBLEM WHAT TO DOWhen this symbolappears and stays on thescreen, it means thebatteries are getting low.This symbol will remainon the displa

Strona 55

MESSAGE PROBLEM WHAT TO DOcleantestarea1. There is dirt, blood, orlint on the test area.2. Your hand or an objectcovered the test areawhile the meter

Strona 56

— How to Test Your Blood 27Special Messages 31Checking the Amount of Blood on the Test Strip 32USING THE METER MEMORY 34Entering the Memory Mode 35Rec

Strona 57 - Error Messages

53MESSAGE PROBLEM WHAT TO DOIf the word contrlappears after a bloodtest, your blood samplewas too small, smeared, oranother drop was addedafter the te

Strona 58 - MESSAGE PROBLEM WHAT TO DO

54MESSAGE PROBLEM WHAT TO DO4. There was not enough blood on the test strip.5. The meter was used in very bright light.6. The check stripprocedure was

Strona 59

MESSAGE PROBLEM WHAT TO DOerror3retestnotenoughbloodretestThe test strip wasremoved during a test orthe test strip was insertedwith blood or controlso

Strona 60

56MESSAGE PROBLEM WHAT TO DO3. Your hand or an objectcovered the test areawhile the meter wasturned on.1. If appearing when youturn your meter on,your

Strona 61

57MESSAGE PROBLEM WHAT TO DO(Example)The check strip testresult is outside theacceptable range (printedon the back of the meter.)The last check strip

Strona 62

58MESSAGE PROBLEM WHAT TO DOcode--The meter requires atest strip code (found on the test strip vial) tooperate. (Note: This isnot an error messagewhen

Strona 63

59TAKING CARE OF YOUR METERYour ONE TOUCH™BASIC™PlusMeter is easy to maintain. These tipswill help you keep it in goodoperating condition: Keep the te

Strona 64

Doing a Daily CheckLook through the small hole in thetest strip holder to make sure there isno lint, dirt, or blood blocking it.If there is anything b

Strona 65

Step 1: Remove the Test StripHolder from the Meter.Hold the meter and place yourthumbs on the two raised dots on thetest strip holder. Press down on t

Strona 66 - YOUR METER

CAUTION: Do not get waterinside the meter. Do notimmerse meter in water.62Step 3: Clean the Test Area.The following cleaning agents willdamage the met

Strona 67 - Cleaning the Meter

SOLVING PROBLEMS 46Test Procedure Messages 46Error Messages 50TAKING CARE OF YOUR METER 59Doing a Daily Check 60Cleaning the Meter 60Replacing the Bat

Strona 68

63Step 4: Replace the Test StripHolder.Hook the bottom of the test stripholder onto the square notch on the meter.Press down on the raised dots of the

Strona 69

64Whenappears flashing on the display, themeter will no longer give results andyou must replace the batteries beforeyou can perform another test.The m

Strona 70

65To replace the batteries:1. Make sure the meter is turned offbefore you remove the batteries.2. Turn the meter over so that thefront of the meter is

Strona 71 - Replacing the Batteries

666. Insert the new AAA batteries intothe battery compartment, beingsure to align the plus (+) andminus (-) signs correctly.To replace the battery com

Strona 72 - (Back of Meter)

67SPECIFICATIONSPower Supply: Two AAA alkalinebatteries.Battery Life: About 18 months whentesting up to two (2) times a day.Result Range: 0–600 mg/dL

Strona 73 - References

68WARRANTYThree-Year Warranty.If, at any time during the first threeyears after purchase, the meter doesnot work for any reason (except forobvious abu

Strona 74 - SPECIFICATIONS

69Before you return your meter, or anyproduct, first call your authorizedLifeScan representative.This warranty is in lieu of all otherwarranties, expr

Strona 75 - WARRANTY

The enclosed LifeScan Blood Glucose Meter, as an electromedicaldevice, has been thoroughly tested and found to comply with theessential requirements o

Strona 76

For further information on LifeScan products, please call your authorized LifeScan representative.© LifeScan, Inc. 2000Milpitas, California 95035, U.S

Strona 77

1PRECAUTIONS AND LIMITATIONSThe following information will beuseful to you and your healthcareprofessional when using the ONE TOUCH™BASIC™Plus Systemt

Strona 78

2Extremes in hematocrit (the amountof red blood cells in the blood) canaffect test results. Very high hemat-ocrits (above 60%) and very lowhematocrits