®companyaHospital Meter Operator’s GuideSSHosp.book Page 1 Thursday, November 14, 2002 3:44 PM

Strona 2

viii Help1Read through the section of the guide specific to the operation you are performing. Refer to the table of contents or index to locate inform

Strona 3

Introduction 1CHAPTER 1IntroductionThe SureStep® Hospital Blood Glucose Meter is one component of LifeScan’s professional blood glucose monitoring sys

Strona 4

Introduction2 Chapter 1 Control Solutions are available to check that the meter and test strips are working together properly.Read the appropriate pac

Strona 5 - Table of Contents

Introduction 3BACKOPENEDBlood Glucose Meter1 800 524-72261 888 353-0800Milpitas, California 95035C22.2 No. 125; Risk Class 2© LifeScan Inc. 1999B1RT

Strona 6

Introduction4 Chapter 1 Display SymbolsThe SureStep Meter uses symbols to guide you through a test and inform you if it detects a problem with the tes

Strona 7

Introduction 5Meter Memory ResultThe meter stores up to 10 blood glucose and control solution test results in memory. When you turn on the meter, the

Strona 8

Introduction6 Chapter 1 Apply Blood Sample to the Test StripThis symbol prompts you to apply sample to the center of the pink test square.Insert Test

Strona 9

Introduction 7Meter is Reading the ResultThe test result appears in approximately 30 seconds. The time may vary depending on temperature and glucose l

Strona 10

Introduction8 Chapter 1 HI ResultThe meter reads HI in virtually all cases when a result is greater than 500 mg/dL (27.8 mmol/L). ◆ IMPORTANTWhen the

Strona 11 - Introduction

Introduction 9Error MessagesWhen the meter detects a problem it displays a specific error message (Er1 through Er6). Refer to Chapter 5 for more infor

Strona 12

SSHosp.book Page 2 Thursday, November 14, 2002 3:44 PM

Strona 13 - 1 888 353-0800

Introduction10 Chapter 1 Replace the Two AAA Alkaline BatteriesWhen this symbol appears, the batteries are low and should be replaced as soon as possi

Strona 14

Introduction 11intensity of the blue is read by the meter—the darker the blue, the higher the blood glucose level.White Tip faces up as the test strip

Strona 15 - Code Number

Introduction12 Chapter 1 Confirmation Dot is read by the meter to determine the sample’s glucose concentration. If it turns completely blue, this indi

Strona 16

Introduction 13• Replace the test strip bottle cap immediately after removing a test strip and close the cap tightly.• Do not use test strips that are

Strona 17 - 0–500 mg/dL (0–27.8 mmol/L)

Introduction14 Chapter 1 Precautions• Unexpected results should be confirmed by a laboratory test.• Various conditions may cause the meter to produce

Strona 18

Introduction 15perform a quality control test. A passing control result indicates the system is working properly. A control result that is out of rang

Strona 19 - Remove the Test Strip

Introduction16 Chapter 1 • Never make significant changes to a medication program or ignore signs and symptoms without consulting a physician.• Failur

Strona 20

Operation 17CHAPTER 2OperationCoding the MeterThe SureStep Meter is calibrated at the factory. Although an internal verification is performed each tim

Strona 21

Operation18 Chapter 2 ▲ CAUTIONMatch the code number on the meter display to the code number on the test strip bottle. If the code numbers do not matc

Strona 22

Operation 192Match the code numbers.The code number on the test strip bottle will range from 1 to 21. If the number on the display matches the code nu

Strona 23

The SureStep Meter is for in vitro diagnostic use for the quantitative measurement of glucose in whole blood samples. It should not be used for the di

Strona 24

Operation20 Chapter 2 the code number on the display matches the code number on the test strip bottle. The code is now set for the package of test str

Strona 25

Operation 21• When you open a new bottle of test strips.• If a patient test has been repeated and the blood glucose results are still lower or higher

Strona 26

Operation22 Chapter 2 • Discard any unused portion 3 months after opening.Write the opened date on the vial when you first open it. • The control rang

Strona 27 - Operation

Operation 23Performing a Quality Control Test1Turn on the meter. Check the display and make sure that all symbol segments appear (see “Display Symbols

Strona 28

Operation24 Chapter 2 ▲ CAUTIONFor accurate results, the code number on the meter must match the code number on the test strip bottle.3Apply a drop of

Strona 29 - Code numbers do not match

Operation 25• Wipe the control solution vial tip with a lint-free cloth or tissue before replacing the cap.4Check the confirmation dot on the back of

Strona 30

Operation26 Chapter 2 ◆ IMPORTANTIf you fail to completely insert the test strip, the test may start; however, you may receive an out-of-range result.

Strona 31 - Glucose Control Solution

Operation 27The control solution result appears in approximately 30 seconds. ◆ IMPORTANTDo not remove the test strip until the countdown is complete.

Strona 32

Operation28 Chapter 2 Control Solution ResultsWhen the control test result falls within the control range, the meter has passed the QC check and can b

Strona 33 - C Button

Operation 29• testing control solutions at high altitude without applying correction factors (read the SureStepPro Glucose Control Solution package in

Strona 34

©2002, 1997 LifeScan, Inc. All rights reserved.The system described herein is covered by one or more of the following US patents: D371,198, D367,109,

Strona 35

Operation30 Chapter 2 Run a linearity solution test as you would a control solution test. Follow the instructions under “Performing a Quality Control

Strona 36

Operation 31Performing a Patient Blood Test1Turn on the meter by pressing the blue power button. Check the display and make sure that all symbol segme

Strona 37 - IMPORTANT

Operation32 Chapter 2 ▲ CAUTIONFor accurate results, the code number on the meter must match the code number on the test strip bottle. 3Obtain a drop

Strona 38

Operation 33SyringeTouch the tip of the syringe to the center of the pink test square. As the blood emerges from the tip of the syringe, it is absorbe

Strona 39 - ▲ CAUTION

Operation34 Chapter 2 PipetteTouch the blood emerging from the tip of a narrow-bore pipette to the center of the pink test square.5Check the white pad

Strona 40

Operation 35The confirmation dot should be completely blue. If white patches or streaks are visible, you have not applied enough blood for an accurate

Strona 41

Operation36 Chapter 2 ▲ CAUTIONIf you fail to completely insert the test strip, the test may start; however, you may receive an inaccurate, low result

Strona 42

Operation 37The test result appears in approximately 30 seconds.◆ IMPORTANTDo not remove the test strip until the countdown is complete. 7Remove the s

Strona 43

Operation38 Chapter 2 Patient ResultsThe meter displays results between 0 and 500 mg/dL (27.8 mmol/L). The meter displays HI for results greater than

Strona 44

Settings and Memory 39CHAPTER 3Settings and MemoryMeter SettingsThe SureStep Meter has three features which are pre-set but may be modified in set-up

Strona 45

iiiTable of ContentsPreface . . . . . . . . . . . . . . . . . . . . . . . . . . . . .vAbout this Guide . . . . . . . . . . . . . . . . . . . . . . .

Strona 46

Settings and Memory40 Chapter 3 Entering Set-up Mode1Turn off the meter by pressing the blue power button.2 Press and hold down the green C button.3 P

Strona 47

Settings and Memory 41quickly through the features, press the blue power button repeatedly or press and hold it to scroll through the settings.The met

Strona 48

Settings and Memory42 Chapter 3 Glucose UnitsThe meter is pre-set to display test results in milligrams per deciliter (mg/dL). You may select to displ

Strona 49 - Settings and Memory

Settings and Memory 43performed correctly. The meter can detect some, but not all errors in the test procedure.• A single short beep confirms each ste

Strona 50

Settings and Memory44 Chapter 3 A three-tone beep sounds when OFF is first displayed. Then it will flash silently until the meter turns off.Exiting Se

Strona 51 - Display of Last Result

Settings and Memory 45symbol appears briefly, showing that there is no test result in the memory.This symbol will also appear after turning on the met

Strona 52

Settings and Memory46 Chapter 3 button. The most recent test number appears (1) followed by the result.Each time the C button is pressed, the next mos

Strona 53 - Beep signals off

Settings and Memory 47If a previous test resulted in an error condition, the error will be stored in memory.Recalling Most Recent Result After Testing

Strona 54

Settings and Memory48 Chapter 3 With the current test result displayed, press and hold the green C button. The previous result will appear. SSHosp.boo

Strona 55

Cleaning and Maintenance 49CHAPTER 4Cleaning and MaintenanceGeneral CareTo keep the meter in good operating condition, you must keep it clean and hand

Strona 56

iv Performing a Quality Control Test . . . . . . . . . 23Control Solution Results . . . . . . . . . . . . . . . . 28Linearity Test . . . . . . . . .

Strona 57 - Current result

Cleaning and Maintenance50 Chapter 4 symbol segments appear (see “Display Symbols” on page 4). Perform a control solution test prior to running a pati

Strona 58

Cleaning and Maintenance 51Cleaning the Outside of the MeterClean the outside of the meter with a cloth moistened with a 10% bleach solution. Follow w

Strona 59 - Maintenance

Cleaning and Maintenance52 Chapter 4 ammonia to clean the test strip holder or lens area because it will damage the parts.1Remove the test strip holde

Strona 60 - Cleaning and Maintenance

Cleaning and Maintenance 53Be sure to thoroughly wipe the gray area on the inside cover. Clean both sides of the base. Follow with a cotton swab or c

Strona 61

Cleaning and Maintenance54 Chapter 4 2Clean the lens area and contact points.Using a cotton swab or a cloth moistened with a 10% bleach solution, wipe

Strona 62

Cleaning and Maintenance 55firmly in place and does not extend beyond the meter. 4Turn on the meter.If the following symbol appears after cleaning, ch

Strona 63

Cleaning and Maintenance56 Chapter 4 was inserted correctly. The test strip holder may be blocking the meter optics. NOTEThis message may also indic

Strona 64

Cleaning and Maintenance 57Replacing the BatteriesThe SureStep Meter comes with two AAA alkaline batteries installed. When this symbol appears the bat

Strona 65 - Turn on the meter

Cleaning and Maintenance58 Chapter 4 2Remove the battery door.Squeeze the tab and lift up. 3Remove the old batteries.Gently tap the meter on the palm

Strona 66

Cleaning and Maintenance 594Insert new batteries.Insert two new AAA alkaline batteries, matching the + end of each battery with the + signs inside the

Strona 67

vPrefaceThe SureStep Meter is a simple, easy-to-use meter for monitoring blood glucose levels.The SureStep Hospital Blood Glucose Monitoring System o

Strona 68

Cleaning and Maintenance60 Chapter 4 SSHosp.book Page 60 Thursday, November 14, 2002 3:44 PM

Strona 69 - AW 052-178-01A Made in U.S.A

Troubleshooting 61CHAPTER 5TroubleshootingThe SureStep Meter displays prompts and error messages to indicate the system has detected a problem with it

Strona 70

Troubleshooting62 Chapter 5 LifeScan Healthcare Professional Line: 1 800 524-7226▲ CAUTIONWhile resolving an error message using the Troubleshooting s

Strona 71 - Troubleshooting

Troubleshooting 63LifeScan Healthcare Professional Line: 1 800 524-7226• Do not use the Color Chart as a replacement for a meter result.If you need he

Strona 72

Troubleshooting64 Chapter 5 LifeScan Healthcare Professional Line: 1 800 524-7226(continued) The meter detected a problem with the color of the test s

Strona 73

Troubleshooting 65LifeScan Healthcare Professional Line: 1 800 524-7226The meter detected a problem with the test strip.1.Repeat test with new test st

Strona 74

Troubleshooting66 Chapter 5 LifeScan Healthcare Professional Line: 1 800 524-7226(continued)The meter detected a problem with the test strip.2.If you

Strona 75

Troubleshooting 67LifeScan Healthcare Professional Line: 1 800 524-7226The meter detected a problem with the position of the test strip.1.Clean meter,

Strona 76

Troubleshooting68 Chapter 5 LifeScan Healthcare Professional Line: 1 800 524-7226The meter detected a temperature problem.1.Repeat test with new test

Strona 77

Troubleshooting 69LifeScan Healthcare Professional Line: 1 800 524-7226The meter detected a problem with the optical system.1.Clean meter, especially

Strona 78

vi About this GuideThe SureStep Hospital Meter Operator’s Guide provides detailed instructions on using and maintaining the SureStep Meter. Although o

Strona 79

Troubleshooting70 Chapter 5 LifeScan Healthcare Professional Line: 1 800 524-7226The meter detected a problem with the electronics.1.Wipe contact poin

Strona 80

Troubleshooting 71LifeScan Healthcare Professional Line: 1 800 524-7226Helpful MessagesSymbol Meaning What To DoTest strip was not fully inserted into

Strona 81 - Helpful Messages

Troubleshooting72 Chapter 5 LifeScan Healthcare Professional Line: 1 800 524-7226If symbol is constant, batteries are low. If symbol is flashing, batt

Strona 82

73SpecificationsThe performance of the system has been evaluated under the conditions listed below. See the SureStepPro Test Strip package insert for

Strona 83 - Specifications

74 hematocrit range: 25–60%dimensions: length 3.5 in (8.9 cm)width 2.4 in (6.1 cm)height 0.8 in (2.0 cm)weight: 3.8 oz with batteriesoperating tempera

Strona 84

75Aapplyingblood 32–34blood, symbol 6control solution 24Bbatterycompartment 3replacing 57–59symbol 10beep signals 42–44bloodapplication 32–34test, pe

Strona 85

76 Ddisplay of last result 41display symbols 4–10Eerror messages 9, 63–70Gglucose control solutionSee control solutionglucose units 42Hhigh result 38I

Strona 86

77symbols, meter 4–10Ttest results 7control solution 27, 28high 8, 38linearity 30patient 37, 38test strip holder 3cleaning 51–53test strip symbolinse

Strona 87

SSHosp.book Page 78 Thursday, November 14, 2002 3:44 PM

Strona 88

AW 050-097-03A©2002 LifeScan, Inc. 11/02SSHosp.book Page 1 Thursday, November 14, 2002 3:44 PM

Strona 89 - ©2002 LifeScan, Inc. 11/02

viiChapter 4, Care and Maintenance, instructs you on how to clean the meter and change the batteries.Chapter 5, Troubleshooting, lists the error code